Pressure Calibration – Field Procedure

Pressure transmitter calibration must always be performed using proper isolation, equalization, and safe handling practices. Most calibration errors occur due to incorrect manifold operation.

Before Starting Calibration

Never attempt calibration without proper isolation.

Step 1 – Safe Isolation

  1. Close HP and LP isolation valves
  2. Open equalizing valve
  3. Ensure transmitter reading drops to zero
  4. Vent process pressure safely

Always equalize before disconnecting any impulse line.

Step 2 – Zero Check

For DP transmitters, zero shift is common after shutdown.

Step 3 – Span Calibration

  1. Close equalizing valve
  2. Apply known pressure using calibration pump
  3. Verify output matches expected mA value
  4. Adjust span if required

Example: If range is 0–10 bar, applying 10 bar should produce 20 mA.

Step 4 – Linearity Check

Always increase and decrease pressure to check hysteresis.

Common Calibration Mistakes

After Calibration – Restoration

  1. Open equalizing valve
  2. Slowly open HP and LP isolation valves
  3. Close equalizing valve
  4. Verify stable reading
  5. Confirm DCS value matches field reading

Sudden valve opening can damage sensor diaphragm.

Field Reality

Many transmitters are replaced unnecessarily. In most cases, incorrect manifold handling or impulse imbalance is the true cause of calibration failure.

← Back to Pressure Hub Related: Zero Shift After Shutdown →