GC Failure Modes – Visual Root Cause Analysis

Most GC problems present themselves on the chromatogram before any alarm appears. This page links common field symptoms to their most likely causes so technicians can troubleshoot logically and quickly.

Symptom → Cause → Action Map

Follow the arrows from observed symptom to root cause, then apply the recommended corrective action.

Symptom Likely Cause Corrective Action Missing Peak Inject valve not switching / timing error Check valve actuation, air supply, timing Baseline Drift / Noise Contamination / moisture / unstable pressure Check filters, dryers, regulators Peak Shift Flow / temperature drift Verify carrier flow, oven control Long Cycle Time Backflush not working / late timing Verify backflush valve and sequence Poor Repeatability Sampling pressure / valve instability Stabilize sample system, check pneumatics

Golden Troubleshooting Rules

Field Tip

If multiple symptoms appear together, suspect a common upstream cause such as contamination, pressure fluctuation, or air supply failure rather than individual component faults.